Operation Manual for LCD Digital Readouts Page 44 of 48
Pr 37 – Programmable Output Polarity. [0 or 1]
Sets the normal state of the programmable output when not activated. The
output is a transistor that conducts to negative.
0 = Normally open.
1 = Normally closed.
Default = 0 (NO)
Pr 38 – Programmable Output Operation. [0, to 3, 0]
Selects the function that the programmable output activates on.
See: CIRCUIT BOARD JUMPER JP5 also.
0 – No Operation.
1 = Drift Monitor
2 = Soft Limits
3 = Go/No Go
Default = 0
Value of Pr 38 Function
0
No Operation
1
Drift (Monitor) Operation
2
Soft Limits
3
Go/No Go